|
1
|
- AMANDA CAMERON
- Forensic Chemist
- Georgia Bureau of Investigation
- Atlanta, Georgia
|
|
2
|
- chro·ma ·tog ·ra ·phy: n. Any of various techniques for the separation
of complex mixtures that rely on the differential affinities of
substances for a gas or liquid mobile medium and for a stationary
absorbing medium through which they pass, such as paper, gelatin, or
magnesia.
- -American Heritage Dictionary of the English Language. 4th ed. 2000.
- From the Greek words “chromatos” meaning “color,” and “graphy” meaning
“writing”
|
|
3
|
- 1903: Russian botanist, Mikhail S. Tswett experimented with separation
of leaf pigments from plants using solvents
- Using an open glass tube packed with calcium carbonate and alumina, he
poured the solvent extract of leaves into the column
- Then, pure solvent was poured through the column, washing the extract
down the column through gravity
- Tswett related the resulting different colored bands to different
compounds present in the original leaf sample
|
|
4
|
- The components of the pigment were separated based on their individual
attraction to the column packing (stationary phase)
- Components that were more attracted to the stationary phase remained in
the column longer while components that were more attracted to the
solvent (mobile phase) moved through the column faster
|
|
5
|
|
|
6
|
- Preliminary technique for qualitative analysisàRf
- “Normal Phase” = relatively POLAR stationary phase, relatively NONPOLAR
mobile phase
- Stationary phase is usually a silica gel coated on glass or plastic
plate
- Mobile phase is solvent or combination of solvents
- Useful when target analyte is fairly polar
|
|
7
|
- Perform extraction to get compound in solution
- “Spot” sample solution, negative control and any standards at origin of
plate, allow solvents to evaporate
- To separate, place plate in closed chamber contacting the bottom edge of
the plate with the mobile phase.
- Mobile advances up the plate through capillary action
- Once solvent has reached top of plate, plate is removed and the results
are visualized using an appropriate technique.
- Examples of visualization methods:
- Sulfuric acid/heat: destructive, leaves charred blots behind
- Ceric stain: destructive, leaves a dark blue blot behind for polar
compounds
- Iodine: semi-destructive, iodine absorbs onto the spots, not permanent
- UV light: non-destructive, long wavelength (background green, spots
dark), short wavelength (plate dark, compounds glow)
|
|
8
|
|
|
9
|
|
|
10
|
|
|
11
|
- Holds solvent bottles
- Can use up to 4 solvents
- SOLVENT CHOICE!
- Useful to employ solvent filters to prevent particulate or impurities
from entering system
|
|
12
|
- Pump draws solvent from solvent bottles through degasser
- Vacuum chamber draws gases dissolved in solvents out into chamber
through semi-permeable membrane of tubing
- Bubbles can cause problems in pressure and flow rate,
chromatography/retention time, as well as problems in the detector
|
|
13
|
- Quaternary
- Good for use in a wide range of research and routine applications
- Either isocratic or gradient analysis
- Flow rates up to 10 mL/min
- Pressure range up to 400 bar at max of 5 mL/min
- Binary
- High pressure mixing
- Extremely reproducible gradients at low flow rates
- Flow rates as low as 50 μL/min, pressures as high as 600 bar in SL
system
|
|
14
|
|
|
15
|
- Isocratic elution
- Utilizes only one solvent
- “Single solvent” can mean combination of solvents, but in a constant
ratio throughout analysis
- Depending on target analyte(s), single solvent may not adequately
separate components
- Gradient elution
- Utilizes a combination of solvents in varying ratios throughout the
analysis
- Mobile phase can range from more polar to less polar (or vice versa) to
ensure adequate separation in a useful amount of time
- Disadvantage: column must re-equilibrate after each run (post-run)
|
|
16
|
|
|
17
|
- Arm retrieves sample vial from tray position and carries it to the
needle-seat.
- The needle pierces the septum of the vial cap and draws up some of the
sample before the arm picks up the vial and returns it to its original
tray position.
- The needle then injects the sample into the solvent flow.
|
|
18
|
|
|
19
|
- Depending on the manufacturer, the column is positioned in a
thermostatted chamber and connected to the injection valve and detector
through a series of capillary tubing.
- Compounds exit the column at a retention time dependent on system
parameters
- Samples continue through the system to the detector
|
|
20
|
- Several column options are available from several manufacturers
- Consider samples to be analyzed
- Consider goals of separation
- Physical features of column:
- Particle size --Surface area
- Column length --Pore size
- Internal diameter --Carbon load
- Bonding type --Particle shape
- www.alltechweb.com (http://www.discoverysciences.com/Productinfo/Technical/posters/MethodDevelopment.ppt#1)
|
|
21
|
|
|
22
|
- C18 (Octadecylsilane)—very nonpolar; retention based on London
interactions with hydrophobic compounds
- Phenyl—nonpolar; retention is a mixed mechanism of hydrophobic and π-π
interactions
- Cyanopropyl—intermediate polarity; retention is a mix of hydophobic,
dipole-dipole and π-π interactions
|
|
23
|
- Specialty silicas
- Sulfono --Fluoro
- Diols --Miscellaneous
- Amino
- Polymers: PS-DVB, polyvinyl alcohol
- Zirconium oxide
|
|
24
|
- Length:
- Longer columns = higher resolution but longer run times
- Shorter columns = short run time, low back pressue, less solvent usage
and waste
- Diameter:
- Narrow columns = higher detector sensitivity, higher pressure required,
less sample can be injected while maintaining good peak shape
- Wide columns = generally used for preparative HPLC
|
|
25
|
- Particle Characteristics
- Shape
- Size
- Pore size
- Surface Area
- Bonding type
- Carbon Load
- All factors have an effect on the number of theoretical plates, and
therefore the efficiency of the column
|
|
26
|
- H = plate height
- λ = particle shape (with regard to the packing)
- dp = particle diameter
- G, ω, and R are constants
- Dm = diffusion coefficient of the mobile phase
- dc = capillary diameter
- df = film thickness
- Ds = diffusion coefficient of the stationary phase.
|
|
27
|
- A = Eddy-diffusion
- B = Longitudinal diffusion
- C = Resistance to mass transfer
- u = Flow rate
|
|
28
|
- Plate Height can be calculated using the equation
- Where N = number of plates and L = length of the column
|
|
29
|
|
|
30
|
|
|
31
|
- Optical Detectors:
- Fixed wavelength
- Fluorescence
- Refractive Index
- Diode Array Detector
- Electrochemical
- Infra-red
- Mass Spectrometry
|
|
32
|
|
|
33
|
|
|
34
|
- A chemical group capable of selective light absorption resulting in the
coloration of certain organic compounds.
|
|
35
|
- Absorbance = -log T =
- -log(I / I0) = εbc where:
- ε = molar absorptivity
- b = length of sample cell (cm)
- c = molar concentration of solute
- T = I/I0 = Intensity of light after passing through
cell/Intensity incident light
|
|
36
|
- Transmittance is measured as a percentage while absorbance becomes
arbitrary > 1.0
|
|
37
|
- Contain low-pressure mercury lamps with an intense emission line at 254
nm
- Other types of lamps are available
- Some have optional additional wavelength settings
- Many organic compounds absorb at 254 nm
- Best for situations when wavelength is rarely varied
|
|
38
|
- Solute is excited with UV radiation and emits radiation at a longer
wavelength
- Can offer great selectivity since excitation and emission wavelengths,
as well as retention times can be used to identify drugs
- Solvent choice is important:
- Mobile phase must neither fluoresce nor absorb at chosen wavelengths
- pH: some drugs only fluoresce in certain ionic forms
|
|
39
|
- Refractive Index of a medium is a measure for how much the speed of
light is reduced inside the medium
- Changes in refractive index (positive or negative) that arise from the
presence of a compound in the eluent are recorded
- Universal detector, but lease sensitive
- Factors must be controlled during separation:
- Temperature
- Eluent composition
- Pressure
|
|
40
|
- Sample travels from column through flow-cell where polychromatic light
is focused on the flow cell and is dispersed onto a chip containing
light sensitive diodes arranged side-by-side
- Each diode registers a well-defined fraction of the UV-spectrum so it
can measure all wavelengths at the same time
|
|
41
|
- Measure the gain or loss of electrons from migrating samples as they
pass between electrodes at a given difference in electrical potential
- Eluents must be electrically conductive (addition of inert electrolytes)
- Most easily used in the oxidative mode (use in reductive mode requires
the removal of dissolved oxygen from the eluent)
|
|
42
|
- Non-destructive technique
- Scans compounds in the IR region and measures the stretching and bending
vibrations of chemical bonds
- Advantage of 3 pieces of data from 1 sample/extraction: retention time,
UV, IR
|
|
43
|
- Presents unique challenge because liquid phase must be converted to gas
phase for normal mass spectral analysis
- Different methods of atmospheric pressure ionization:
- Electrospray ionization (ESI) - voltage applied to the solvent - as
solvent evaporates electrical field at the surface of the drop
increases due to the decreasing radius of the droplet
- Atmospheric pressure chemical ionization (APCI) - heat evaporates the
solvent and sample charged via charged probe.
- API: need volatile mobile-phase modifiers in the chromatographic
separation
|
|
44
|
- Additional component that allows for collection of individual components
from a multi-component sample
- Can be set to collect certain peak from multiple injections of sample in
one vial
|
|
45
|
- Pharmaceutical (drug development, research, QA)
- Clinical (drug monitoring, drugs of abuse, disease markers)
- Environmental analysis
- Food & beverage
- Polymers & plastics
- Forensics!!
|
|
46
|
- GBI: (primarily) Agilent 1100 HPLC, Zorbax XDB C18 4.6 x 150 mm, 5 μm
column, DAD detector
|
|
47
|
|
|
48
|
|
|
49
|
|
|
50
|
|
|
51
|
- Mobile Phase pH
- Injection Valve
- Metering Device
- Needle and Needle Seat
- Column Compartment
- Capillary/Peek Tubing and fittings
- Article on carryover: http://www.forumsci.co.il/HPLC/Carryover_problems.pdf
|
|
52
|
- Reproducibility
- Metered injection
- Stable solvents
- Temperature control
- Peak resolution
- Relatively short analysis time
- Ease of sample prep
- Ease of calculation
- Ease of explanation
|
|
53
|
- Standard Addition
- Internal Standard
- External Standard
|
|
54
|
- Useful when sample is in a complex matrix
- Multiple aliquots of unknown sample are prepared.
- One sample is analyzed straight for the target
- Other samples are spiked with precise but varying amounts of the target
- Calibration curve of response vs. concentration is produced to directly
determine the unknown
- Disadvantages:
- Difficult to explain --Time-consuming preparation
- Requires greater volume of sample
- Associated uncertainty is large
|
|
55
|
- Tends to yield the most accurate and precise results
- A known amount of a component that is chemically similar and has a
retention time close to the target analyte, but that is not present in
the sample, is added to the target and a standard solution
- Ratios of peak height or area of the target to the internal standard
gives quantitation data
|
|
56
|
- Concunk = Areaint. std.in known x Areaunknown
x Concknown
- Areaint.
std.in unknown Areaknown
|
|
57
|
- Reference standards of varying amounts encompassing the expected range
of sample concentration are used to generate a calibration curve
- Simplest quantitation technique
- Easy sample prep
- Easy to plot and explain
- Requires consistency in analysis (consistent, reliable injections)
- Requires additional QA/QC procedures
|
|
58
|
|
|
59
|
|
|
60
|
|
|
61
|
|
|
62
|
- Once calibration curve is established, it is considered valid until
something invalidates it
- **Need additional QA/QC measures to ensure instrument is still working
properly and to ensure that quantitation procedure is followed properly
|
|
63
|
- Analyzed with a negative control (ethanol) at the beginning and end of
each sequence
- Also analyzed before and after each scientist to bracket individual data
- Considered valid if within 5% of known concentration
|
|
64
|
- Standard of known purity
- Quantitatively weighed and analyzed along with samples
- Calculate the % recovery of the SCS for each scientist
- %Purity = conc (mg/mL) * Vol (mL) * 100
- weight (mg)
- %Recovery = . % Purity . *100
- Actual Purity
|
|
65
|
- GBI reports at a 95% confidence interval (corresponds to 1.96 σ)
- For sample/dup, meas. unc. for process = 4.25%
- For 5 bag quant: Meas Unc =( TN-1 * SD)/ √N
- where N = number of pkgs tested, TN-1 = T value at 95%
confidence and N-1 degrees of freedom, SD = Standard deviation of
samples
|
|
66
|
|
|
67
|
- Convenient separation of wide range of sample types
- Excellent resolving power
- Speed of analysis
- Low detection levels for weak samples
|
|
68
|
|